English

Description of a Differential Setup for Relaxation Microcalorimetry

Instrumentation and Detectors 2013-03-20 v1 Materials Science

Abstract

A specific heat measuring instrument, with the capacity for the application of magnetic fields up to 9 Tesla, resorting to microcalorimetry chips from the company Xensor Integration, has been successfully assembled and its functioning specifications are reported in the current paper. With this instrument is it possible to perform specific heat measurements with applied magnetic fields up to 9 Tesla in milligram samples. This offers our group the possibility to calculate the actual adiabatic temperature change of a material, as well as providing reliable and precise information on any phase transition that may be influenced by the application of a magnetic field.

Keywords

Cite

@article{arxiv.1303.4544,
  title  = {Description of a Differential Setup for Relaxation Microcalorimetry},
  author = {J. V. Leitao and P. van Dommelen and F. Naastepad and E. Brück},
  journal= {arXiv preprint arXiv:1303.4544},
  year   = {2013}
}

Comments

Previously published in the proceedings of Thermag V and currently submitted for a Special Issue of the International Journal of Refrigeration

R2 v1 2026-06-21T23:44:20.090Z