English

Charge collection efficiency in back-illuminated Charge-Coupled Devices

Instrumentation and Detectors 2021-06-16 v1 High Energy Physics - Experiment

Abstract

Low noise CCDs fully-depleted up to 675 micrometers have been identified as a unique tool for Dark Matter searches and low energy neutrino physics. The charge collection efficiency (CCE) for these detectors is a critical parameter for the performance of future experiments. We present here a new technique to characterize CCE in back-illuminated CCDs based on soft X-rays. This technique is used to characterize two different detector designs. The results demonstrate the importance of the backside processing for detection near threshold, showing that a recombination layer of a few microns significantly distorts the low energy spectrum. The studies demonstrate that the region of partial charge collection can be reduced to less than 1 micrometer thickness with adequate backside processing.

Keywords

Cite

@article{arxiv.2007.04201,
  title  = {Charge collection efficiency in back-illuminated Charge-Coupled Devices},
  author = {Guillermo Fernandez-Moroni and Kevin Andersson and Ana Botti and Juan Estrada and Dario Rodrigues and Javier Tiffenberg},
  journal= {arXiv preprint arXiv:2007.04201},
  year   = {2021}
}
R2 v1 2026-06-23T16:57:20.737Z