Widerange $T_\alpha$-$Q_\alpha$ formula for real time application
Instrumentation and Detectors
2022-06-01 v3 Nuclear Experiment
Abstract
A new wide-range formula to operate with the new DGFRS2 analog spectrometer installed at the DC-280 cyclotron facility is presented. The main goal of this formula application algorithm is to search the optimal time correlation recoil-alpha parameter directly during the execution of the C++ acquisition code. Note that the spectrometer operates together with the 48x128 strip DSSD (Double Side Strip Detector) detector and low-pressure pentane-filled gaseous detector. Comparison with others known formulas is performed for Z=119,120 nuclei.
Cite
@article{arxiv.2205.11810,
title = {Widerange $T_\alpha$-$Q_\alpha$ formula for real time application},
author = {D. Ibadullayev and Yu. S. Tsyganov},
journal= {arXiv preprint arXiv:2205.11810},
year = {2022}
}
Comments
will replace with more correct data (Table1) in a nearest future