English

Waveform Timing Performance of a 5 GS/s Fast Pulse Sampling Module with DRS4

Instrumentation and Detectors 2015-10-28 v1 High Energy Physics - Experiment

Abstract

We first clarify timing issues of non-uniform sampling intervals regarding a 5 GS/s fast pulse sampling module with DRS4. Calibration strategy is proposed, and as a result, the waveform timing performance is improved to be below 10 ps RMS. We then further evaluate waveform timing performance of the module by comparing with a 10 GS/s oscilloscope in a setup with plastic scintillators and fast PMTs. Different waveform timing algorithms are employed for analysis, and the module shows comparable timing performance with that of the oscilloscope.

Keywords

Cite

@article{arxiv.1501.00651,
  title  = {Waveform Timing Performance of a 5 GS/s Fast Pulse Sampling Module with DRS4},
  author = {Jinhong Wang and Shubin Liu and Qi An},
  journal= {arXiv preprint arXiv:1501.00651},
  year   = {2015}
}
R2 v1 2026-06-22T07:50:14.010Z