Inductively Coupled Plasma Mass Spectrometry (ICP-MS) allows for rapid, high-sensitivity determination of trace impurities, notably the primordial radioisotopes 238U and 232Th, in candidate materials for low-background rare-event search experiments. We describe the setup and characterisation of a dedicated low-background screening facility at University College London where we operate an Agilent 7900 ICP-MS. The impact of reagent and carrier gas purity is evaluated and we show that twice-distilled ROMIL-SpA-grade nitric acid and zero-grade Ar gas delivers similar sensitivity to ROMIL-UpA-grade acid and research grade gas. A straightforward procedure for sample digestion and analysis of materials with U/Th concentrations down to 10 ppt g/g is presented. This includes the use of 233U and 230Th spikes to correct for signal loss from a range of sources and verification of 238U and 232Th recovery through digestion and analysis of a certified reference material with a complex sample matrix. Finally, we demonstrate assays and present results from two sample preparation and assay methods: a high-sensitivity measurement of ultra-pure Ti using open digestion techniques, and a closed vessel microwave digestion of a nickel-chromium-alloy using a multi-acid mixture.
@article{arxiv.1708.08860,
title = {Ultra-low background mass spectrometry for rare-event searches},
author = {J. Dobson and C. Ghag and L. Manenti},
journal= {arXiv preprint arXiv:1708.08860},
year = {2018}
}