Total-Reflection High-Energy Positron Diffractometer at NEPOMUC -- Instrumentation, Simulation and First Measurements

Instrumentation and Detectors 2021-11-05 v1

Abstract

We report the instrumentation of a new positron diffractometer that is connected to the high-intensity positron source NEPOMUC. Crucial elements for the adaption of the positron beam are presented, which include the magnetic field termination, the optional transmission-type remoderator for brightness enhancement and the electrostatic system for acceleration and beam optics. The positron trajectories of the remoderated and the twofold remoderated beam have been simulated to optimize the system, i.e. to obtain a coherent beam of small diameter. Within a first beamtime we tuned the system and characterized the direct beam. For the twofold remoderated beam of 10keV energy, we experimentally observe a beam diameter of d < 1.3mm, which agrees well with the simulation.

Keywords

Cite

@article{arxiv.2107.03052,
  title  = {Total-Reflection High-Energy Positron Diffractometer at NEPOMUC -- Instrumentation, Simulation and First Measurements},
  author = {Matthias Dodenhöft and Sebastian Vohburger and Christoph Hugenschmidt},
  journal= {arXiv preprint arXiv:2107.03052},
  year   = {2021}
}

Comments

10 pages, 6 figures