English

Tomographic Imaging by a Si/CdTe Compton Camera for In-111 and I-131 Radionuclides

Instrumentation and Detectors 2021-08-24 v1 High Energy Astrophysical Phenomena Instrumentation and Methods for Astrophysics Medical Physics

Abstract

Tomographic imaging with radionuclides commonly used in nuclear medicine, such as 111^{111}In (171 and 245 keV) and 131^{131}I (364 keV), is in high demand for medical applications and small animal imaging. The Si/CdTe Compton camera with its high angular and high energy resolutions is an especially promising detector to extend the energy coverage for imaging to the range that covers gamma-ray emitted from these radionuclides. Here, we take the first steps towards short-distance imaging by conducting experiments using three-dimensional phantoms composed of multiple sphere-like solutions of 111^{111}In and 131^{131}I with a diameter of 2.7 mm, placed at a distance of 41 mm. Using simple back-projection methods, the positions of the sources are reproduced with a spatial resolution of 11.5 mm and 9.0 mm (FWHM) for 111^{111}In and 131^{131}I, respectively. We found that a LM-MLEM method gives a better resolution of 4.0 mm and 2.7 mm (FWHM). We resolve source positions of a tetrahedron structure with a source-to-source separation of 28 mm. These findings demonstrate that Compton Cameras have the potential of close-distance imaging of radioisotopes distributions in the energy range below 400 keV.

Keywords

Cite

@article{arxiv.2108.10149,
  title  = {Tomographic Imaging by a Si/CdTe Compton Camera for In-111 and I-131 Radionuclides},
  author = {Goro Yabu and Hiroki Yoneda and Tadashi Orita and Shin'ichiro Takeda and Pietro Caradonna and Tadayuki Takahashi and Shin Watanabe and Fumiki Moriyama},
  journal= {arXiv preprint arXiv:2108.10149},
  year   = {2021}
}