English

Thermal Kinetic Inductance Detectors for millimeter-wave detection

Instrumentation and Detectors 2020-07-31 v2 Instrumentation and Methods for Astrophysics

Abstract

Thermal Kinetic Inductance Detectors (TKIDs) combine the excellent noise performance of traditional bolometers with a radio frequency multiplexing architecture that enables the large detector counts needed for the next generation of millimeter-wave instruments. In this paper, we first discuss the expected noise sources in TKIDs and derive the limits where the phonon noise contribution dominates over the other detector noise terms: generation-recombination, amplifier, and two-level system (TLS) noise. Second, we characterize aluminum TKIDs in a dark environment. We present measurements of TKID resonators with quality factors of about 10510^5 at 80 mK. We also discuss the bolometer thermal conductance, heat capacity, and time constants. These were measured by the use of a resistor on the thermal island to excite the bolometers. These dark aluminum TKIDs demonstrate a noise equivalent power NEP = 2×1017W/Hz2 \times 10^{-17} \mathrm{W}/\mathrm{\sqrt{Hz}} , with a 1/f1/f knee at 0.1 Hz, which provides background noise limited performance for ground-based telescopes observing at 150 GHz.

Keywords

Cite

@article{arxiv.2001.08887,
  title  = {Thermal Kinetic Inductance Detectors for millimeter-wave detection},
  author = {Albert Wandui and Jamie Bock and Clifford Frez and Matthew Hollister and Lorenzo Minutolo and Hien Nguyen and Bryan Steinbach and Anthony Turner and Jonas Zmuidzinas and Roger O'Brient},
  journal= {arXiv preprint arXiv:2001.08887},
  year   = {2020}
}

Comments

15 pages, 12 figures

R2 v1 2026-06-23T13:19:35.587Z