English

The KATRIN Pre-Spectrometer at reduced Filter Energy

Instrumentation and Detectors 2015-06-04 v1

Abstract

The KArlsruhe TRItium Neutrino experiment, KATRIN, will determine the mass of the electron neutrino with a sensitivity of 0.2 eV (90% C.L.) via a measurement of the beta-spectrum of gaseous tritium near its endpoint of E_0 =18.57 keV. An ultra-low background of about b = 10 mHz is among the requirements to reach this sensitivity. In the KATRIN main beam-line two spectrometers of MAC-E filter type are used in a tandem configuration. This setup, however, produces a Penning trap which could lead to increased background. We have performed test measurements showing that the filter energy of the pre-spectrometer can be reduced by several keV in order to diminish this trap. These measurements were analyzed with the help of a complex computer simulation, modeling multiple electron reflections both from the detector and the photoelectric electron source used in our test setup.

Keywords

Cite

@article{arxiv.1203.2444,
  title  = {The KATRIN Pre-Spectrometer at reduced Filter Energy},
  author = {M. Prall and P. Renschler and F. Glück and A. Beglarian and H. Bichsel and L. Bornschein and Z. Chaoui and G. Drexlin and F. Fränkle and S. Görhardt and S. Mertens and M. Steidl and Th. Thümmler and S. Wüstling and C. Weinheimer and S. Zadorozhny},
  journal= {arXiv preprint arXiv:1203.2444},
  year   = {2015}
}

Comments

22 pages, 12 figures

R2 v1 2026-06-21T20:32:31.958Z