The High Voltage distribution system of the ATLAS Tile Calorimeter and its performance during data taking

Instrumentation and Detectors 2018-08-17 v3 High Energy Physics - Experiment

Abstract

This article documents the characteristics of the high voltage (HV) system of the hadronic calorimeter TileCal of the ATLAS experiment. Such a system is suitable to supply reliable power distribution into particles physics detectors using a large number of PhotoMultiplier Tubes (PMTs). Measurements performed during the 2015 and 2016 data taking periods of the ATLAS detector show that its performance, in terms of stability and noise, fits the specifications. In particular, almost all the PMTs show a voltage instability smaller than 0.5 V corresponding to a gain stability better than 0.5%. A small amount of channels was found not working correctly. To diagnose the origin of such defects, the results of the HV measurements were compared to those obtained using a Laser system. The analysis shows that less than 0.2% of the about 10 thousand HV channels were malfunctioning.

Keywords

Cite

@article{arxiv.1804.05036,
  title  = {The High Voltage distribution system of the ATLAS Tile Calorimeter and its performance during data taking},
  author = {D. Calvet and S. Calvet and R. Chadelas and D. Cinca and P. Grenier and P. Gris and P. Lafarguette and D. Lambert and M. Marjanović and L. F. Oleiro Seabra and F. M. Pedro Martins and J. B. Pena Madeira Gouveia De Campos and S. M. Romano Saez and P. Rosnet and C. Santoni and L. Valéry and F. Vazeille},