English

The Application of VMM3A Readout for Multi-Grid Neutron Detectors

Instrumentation and Detectors 2025-10-03 v1

Abstract

The T-REX neutron spectrometer at the European Spallation Source will use Multi-Grid Technology, which is a voxelised proportional counter relying on\mathrm{^{10}B_{4}C} coatings to detect the scattered neutrons. Measurements of the position dependence of pulse-height and relative detection efficiency of a Multi-Grid prototype of the T-REX spectrometer are presented for two different schemes of signal-processing electronics based on the VMM3A ASIC and CREMAT technology. These measurements, intended to test the suitability of VMM3A for readout of the T-REX Multi-Grid, are compared with Monte Carlo simulations based on the Garfield++ and Geant4 tool kits.

Keywords

Cite

@article{arxiv.2510.01981,
  title  = {The Application of VMM3A Readout for Multi-Grid Neutron Detectors},
  author = {A. Backis and F. Piscitelli and D. Pfeiffer and J. R. M. Annand and M. Aouane and K. G. Fissum and K. Livingston and G. Mauri and D. Raspino},
  journal= {arXiv preprint arXiv:2510.01981},
  year   = {2025}
}

Comments

14 pages, 9 figures