English

SNO+ scintillator cocktail studies using an ${}^{90}$Y source

Instrumentation and Detectors 2016-05-11 v2

Abstract

We present the design of an 90{}^{90}Y calibration source and its manufacturing procedure, that has been implemented in the University of Sussex radioactive laboratory. The radioactive source was first tested at the University of Sussex using a small scintillator cocktail sample. Further measurements were performed at the University of Pennsylvania using a larger volume of the scintillator cocktail. The results of both studies are presented and discussed.

Cite

@article{arxiv.1605.00134,
  title  = {SNO+ scintillator cocktail studies using an ${}^{90}$Y source},
  author = {Evelina Arushanova and Jeanne R. Wilson},
  journal= {arXiv preprint arXiv:1605.00134},
  year   = {2016}
}

Comments

Poster presented at NuPhys2015 (London, 16-18 December 2015). 4 pages, LaTeX, 5 png figures