English

Semidefinite relaxations in optimal experiment design with application to substrate injection for hyperpolarized MRI

Systems and Control 2016-03-23 v2 Optimization and Control

Abstract

We consider the problem of optimal input design for estimating uncertain parameters in a discrete-time linear state space model, subject to simultaneous amplitude and l1/l2-norm constraints on the admissible inputs. We formulate this problem as the maximization of a (non-concave) quadratic function over the space of inputs, and use semidefinite relaxation techniques to efficiently find the global solution or to provide an upper bound. This investigation is motivated by a problem in medical imaging, specifically designing a substrate injection profile for in vivo metabolic parameter mapping using magnetic resonance imaging (MRI) with hyperpolarized carbon-13 pyruvate. In the l2-norm-constrained case, we show that the relaxation is tight, allowing us to efficiently compute a globally optimal injection profile. In the l1-norm-constrained case the relaxation is no longer tight, but can be used to prove that the boxcar injection currently used in practice achieves at least 98.7% of the global optimum.

Keywords

Cite

@article{arxiv.1510.00455,
  title  = {Semidefinite relaxations in optimal experiment design with application to substrate injection for hyperpolarized MRI},
  author = {John Maidens and Murat Arcak},
  journal= {arXiv preprint arXiv:1510.00455},
  year   = {2016}
}

Comments

6 pages, Accepted to ACC 2016

R2 v1 2026-06-22T11:10:56.134Z