English

Primary Ionization and Particle Identification with Straw Tube Detectors

Instrumentation and Detectors 2025-06-10 v3 High Energy Physics - Experiment

Abstract

The charged particles are tracked in the high energy physics detectors to provide the information of their properties. One of the tracking detector is straw tube detector that has been used by many experiments. The motivation behind the current work is to study the primary ionization and particle identification using straw tube detectors. Additionally, we report the decay of 60^{60}CO for the study of gamma peaks since these are used in cobalt therapy, that is beneficial for cancer/tumor treatment. The various studies like primary ionization, spatial co-ordinate distributions in the different gas mixtures, transition radiation, drift velocities of electrons and diffusion coefficients using different xenon-based gas mixtures have been obtained. These studies have been done after the optimization of xenon-based gas mixtures for a deeper understanding. The gas mixture that shows maximum transition radiation among the xenon-based gas mixtures was found to be Xe:He:CH4Xe:He:CH_{4} :: 30:55:15. The gas mixture that posses maximum primary ionization has been observed to be Xe:CO2Xe:CO_{2} :: 70:30. Finally, the simulations have been carried out for the particle identification in the straw tube detectors with different particles i.e., muons, pions and kaons.

Keywords

Cite

@article{arxiv.2409.10879,
  title  = {Primary Ionization and Particle Identification with Straw Tube Detectors},
  author = {R. Kanishka},
  journal= {arXiv preprint arXiv:2409.10879},
  year   = {2025}
}

Comments

27 pages, 20 figures