English

Performance of the PrimEx Electromagnetic Calorimeter

Instrumentation and Detectors 2019-08-14 v1

Abstract

We report the design and performance of the hybrid electromagnetic calorimeter consisting of 1152 PbWO4PbWO_4 crystals and 576 lead glass blocks for the PrimEx experiment at the Jefferson Laboratory. The detector was built for high precision measurement of the neutral pion lifetime via the Primakoff effect. Calorimeter installation and commissioning was completed with the first physics run in fall of 2004. We present the energy and position resolution of the calorimeter. Obtained π0\pi^0 mass resolution of 1.3MeV/c21.3 \mathrm{MeV/c^2} and its production angle resolution of 0.34mrad0.34 \mathrm{mrad} demonstrate the ability of the experiment to extract the π0\pi^0 lifetime on one percent level.

Keywords

Cite

@article{arxiv.physics/0609201,
  title  = {Performance of the PrimEx Electromagnetic Calorimeter},
  author = {M. Kubantsev and I. Larin and A. Gasparyan},
  journal= {arXiv preprint arXiv:physics/0609201},
  year   = {2019}
}

Comments

8 pages, 7 figures, talk on the CALOR06 Conference, Chicago

R2 v1 2026-07-22T19:12:49.253Z