Although deep learning (DL) has received much attention in accelerated MRI, recent studies suggest small perturbations may lead to instabilities in DL-based reconstructions, leading to concern for their clinical application. However, these works focus on single-coil acquisitions, which is not practical. We investigate instabilities caused by small adversarial attacks for multi-coil acquisitions. Our results suggest that, parallel imaging and multi-coil CS exhibit considerable instabilities against small adversarial perturbations.

Keywords

Cite

@article{arxiv.2102.13066,
  title  = {On Instabilities of Conventional Multi-Coil MRI Reconstruction to Small Adverserial Perturbations},
  author = {Chi Zhang and Jinghan Jia and Burhaneddin Yaman and Steen Moeller and Sijia Liu and Mingyi Hong and Mehmet Akçakaya},
  journal= {arXiv preprint arXiv:2102.13066},
  year   = {2021}
}

Comments

To appear in Proceedings of the 29th Annual Meeting of ISMRM, 2021