Charge-coupled devices (CCDs) are a leading technology in direct dark matter searches because of their eV-scale energy threshold and high spatial resolution. The sensitivity of future CCD experiments could be enhanced by distinguishing nuclear recoil signals from electronic recoil backgrounds in the CCD silicon target. We present a technique for event-by-event identification of nuclear recoils based on the spatial correlation between the primary ionization event and the lattice defect left behind by the recoiling atom, later identified as a localized excess of leakage current under thermal stimulation. By irradiating a CCD with an 241Am9Be neutron source, we demonstrate >93% identification efficiency for nuclear recoils with energies >150 keV, where the ionization events were confirmed to be nuclear recoils from topology. The technique remains fully efficient down to 90 keV, decreasing to 50% at 8 keV, and reaching (6±2)% at 1.5--3.5 keV. Irradiation with a 24Na γ-ray source shows no evidence of defect generation by electronic recoils, with the fraction of electronic recoils with energies <85 keV that are spatially correlated with defects <0.1%.
@article{arxiv.2309.07869,
title = {Nuclear Recoil Identification in a Scientific Charge-Coupled Device},
author = {K. J. McGuire and A. E. Chavarria and N. Castello-Mor and S. Lee and B. Kilminster and R. Vilar and A. Alvarez and J. Jung and J. Cuevas-Zepeda and C. De Dominicis and R. Gaïor and L. Iddir and A. Letessier-Selvon and H. Lin and S. Munagavalasa and D. Norcini and S. Paul and P. Privitera and R. Smida and M. Traina and R. Yajur and J-P. Zopounidis},
journal= {arXiv preprint arXiv:2309.07869},
year = {2024}
}