English

Multi-Frame Energy-Selective Imaging System for Fast-Neutron Radiography

Nuclear Experiment 2015-05-13 v1

Abstract

A new instrument for high resolution imaging of fast-neutrons is presented here. It is designed for energy selective radiography in nanosecond-pulsed broad-energy (1 - 10 MeV) neutron beams. The device presented here is based on hydrogenous scintillator screens and single- or multiple-gated intensified camera systems (ICCD). A key element is a newly developed optical amplifier which generates sufficient light for the high-speed intensified camera system, even from such faint light sources as fast plastic and liquid scintillators. Utilizing the Time-of-Flight (TOF) method, the detector incorporating the above components is capable of simultaneously taking up to 8 images, each at a different neutron energy.

Keywords

Cite

@article{arxiv.0905.3975,
  title  = {Multi-Frame Energy-Selective Imaging System for Fast-Neutron Radiography},
  author = {Volker Dangendorf and Doron Bar and Benjamin Bromberger and Gennady Feldman and Mark B. Goldberg and Ronald Lauck and Ilan Mor and Kai Tittelmeier and David Vartsky and Mathias Weierganz},
  journal= {arXiv preprint arXiv:0905.3975},
  year   = {2015}
}

Comments

6 pages; to be pubished in IEEE TNS, 2009

R2 v1 2026-06-21T13:05:36.396Z