English

Micropascal-sensitivity ultrasound sensors based on optical microcavities

Optics 2022-11-16 v1 Applied Physics

Abstract

Whispering gallery mode (WGM) microcavities have been widely used for high-sensitivity ultrasound detection, due to their optical and mechanical resonances enhanced sensitivity. The ultrasound sensitivity of the cavity optomechanical system is fundamentally limited by the thermal noise. In this work, we theoretically and experimentally investigate the thermal-noise-limited sensitivity of a WGM microdisk ultrasound sensor, and optimize the sensitivity by varying the radius and thickness of the microdisk, as well as using a trench structure around the disk. Using a microdisk with a radius of 300 um and thickness of 2 um, a peak sensitivity of 1.18 uPa Hz^{-1/2} is achieved at 82.6 kHz, which is to our knowledge the record sensitivity among the cavity optomechanical ultrasound sensors. Such high sensitivity can improve the detection range of air-coupled ultrasound sensing technology.

Keywords

Cite

@article{arxiv.2211.07965,
  title  = {Micropascal-sensitivity ultrasound sensors based on optical microcavities},
  author = {Hao Yang and Xuening Cao and Zhi-Gang Hu and Yimeng Gao and Yuechen Lei and Min Wang and Zhanchun Zuo and Xiulai Xu and Bei-Bei Li},
  journal= {arXiv preprint arXiv:2211.07965},
  year   = {2022}
}
R2 v1 2026-06-28T05:55:44.066Z