English

Mass-sensing BioCD Protein Array towards Clinical Application: Prostate Specific Antigen Detection in Patient Sera

Biological Physics 2009-07-27 v1

Abstract

Mass-sensing biosensor arrays for protein detection require no fluorophores or enzyme labels. However, few mass biosensor protein arrays have demonstrated successful application in high background samples, such as serum. In this paper, we test the BioCD as a mass biosensor based on optical interferometry of antibodies covalently attached through Schiff-base reduction. We use the BioCD to detect prostate specific antigen (PSA, a biomarker of prostate cancer) in patient sera in a 96-well anti-PSA microarray. We have attained a 4 ng/ml detection limit in full serum and have measured PSA concentrations in three patient sera.

Cite

@article{arxiv.0907.4310,
  title  = {Mass-sensing BioCD Protein Array towards Clinical Application: Prostate Specific Antigen Detection in Patient Sera},
  author = {Xuefeng Wang and Ming Zhao and David D Nolte and Timothy L Ratliff},
  journal= {arXiv preprint arXiv:0907.4310},
  year   = {2009}
}
R2 v1 2026-06-21T13:28:43.054Z