Lineshape-asymmetry-caused shift in atomic interferometers
Abstract
We investigate the shift caused by asymmetry of spectroscopic lineshape in atomic interferometers, which has not previously been discussed in the scientific literature. This asymmetry arises because laser field is frequency-chirped not only during the free-evolution intervals of atoms, but also during the Ramsey pulses. As a result, the effective detuning from the working atomic transition during the pulses also depends on the chirping rate, which, in turn, leads to the lineshape-asymmetry-caused shift (LACS). It is shown that this shift has an inverse cubic dependence of on the duration of the interval between the Ramsey pulses , which markedly contrasts with the dependence typical in atomic interferometry. Therefore, the metrological importance of this shift substantially increases for compact atomic interferometers with a short baseline. For example, for interferometers-gravimeters using two-photon transitions in rubidium atoms, at ~ms we estimate the LACS shift and its variations at the level of 0.1-1~mGal, while for ~s this can reach a value of 0.1-1~Gal.
Keywords
Cite
@article{arxiv.2512.18476,
title = {Lineshape-asymmetry-caused shift in atomic interferometers},
author = {V. I. Yudin and O. N. Prudnikov and A. V. Taichenachev and M. Yu. Basalaev and D. N. Kapusta and A. N. Goncharov and M. D. Radchenko and V. G. Pal'chikov and L. Zhou and M. S. Zhan},
journal= {arXiv preprint arXiv:2512.18476},
year = {2026}
}
Comments
5 pages, 4 figures