English

Increasing the resolution of transmission electron microscopy by computational ghost imaging

Instrumentation and Detectors 2024-11-20 v1 Optics

Abstract

By means of numerical simulations, we demonstrate the innovative use of computational ghost imaging in transmission electron microscopy to retrieve images with a resolution that overcomes the limitations imposed by coherent aberrations. The method requires measuring the intensity on a single pixel detector with a series of structured illuminations. The success of the technique is improved if the probes are made to resemble the sample and the patterns cover the area of interest evenly. By using a simple 8 electrode device as a specific example, a 2-fold increase in resolution beyond the aberration limit is demonstrated to be possible under realistic experimental conditions.

Keywords

Cite

@article{arxiv.2401.01898,
  title  = {Increasing the resolution of transmission electron microscopy by computational ghost imaging},
  author = {P. Rosi and L. Viani and E. Rotunno and S. Frabboni and A. H. Tavabi and R. E. Dunin-Borkowski and A. Roncaglia and V. Grillo},
  journal= {arXiv preprint arXiv:2401.01898},
  year   = {2024}
}

Comments

8 pages, 3 figures

R2 v1 2026-06-28T14:08:05.569Z