Improved method for measuring low concentration radium and its application to the Super-Kamiokande Gadolinium project

Instrumentation and Detectors 2021-10-28 v1 Instrumentation and Methods for Astrophysics High Energy Physics - Experiment

Abstract

Chemical extraction using a molecular recognition resin named "Empore Radium Rad Disk" was developed to improve sensitivity for the low concentration of radium (Ra). Compared with the previous method, the extraction process speed was improved by a factor of three and the recovery rate for 226^{226}Ra was also improved from 81±\pm4% to >>99.9%. The sensitivity on the 101^{-1} mBq level was achieved using a high purity germanium detector. This improved method was applied to determine 226^{226}Ra in Gd2_2(SO4_4)3_3{\cdot}8H2_2O which will be used in the Super-Kamiokande Gadolinium project. The improvement and measurement results are reported in this paper.

Keywords

Cite

@article{arxiv.2006.09664,
  title  = {Improved method for measuring low concentration radium and its application to the Super-Kamiokande Gadolinium project},
  author = {S. Ito and K. Ichimura and Y. Takaku and K. Abe and M. Harada and M. Ikeda and H. Ito and Y. Kishimoto and Y. Nakajima and T. Okada and H. Sekiya},
  journal= {arXiv preprint arXiv:2006.09664},
  year   = {2021}
}

Comments

8 pages, 3 figures