English

First Point-Spread Function and X-Ray Phase-Contrast Imaging Results With an 88-mm Diameter Single Crystal

Instrumentation and Detectors 2018-08-15 v1 Accelerator Physics

Abstract

In this study, we report initial demonstrations of the use of single crystals in indirect x-ray imaging with a benchtop implementation of propagation-based (PB) x-ray phase contrast imaging. Based on single Gaussian peak fits to the x-ray images, we observed a four times smaller system point-spread function (PSF) with the 50-{\mu}m thick single crystal scintillators than with the reference polycrystalline phosphor/scintillator. Fiber-optic plate depth-of-focus and Al reflective-coating aspects are also elucidated. Guided by the results from the 25-mm diameter crystal samples, we report additionally the first results with a unique 88-mm diameter single crystal bonded to a fiber optic plate and coupled to the large format CCD. Both PSF and x-ray phase contrast imaging data are quantified and presented.

Keywords

Cite

@article{arxiv.1807.02170,
  title  = {First Point-Spread Function and X-Ray Phase-Contrast Imaging Results With an 88-mm Diameter Single Crystal},
  author = {A. H. Lumpkin and A. B. Garson and M. A. Anastasio},
  journal= {arXiv preprint arXiv:1807.02170},
  year   = {2018}
}

Comments

17 pp. Version accepted by Rev.Sci.Inst

R2 v1 2026-06-23T02:52:21.119Z