Experimental Analysis of PandaX-4T Cryogenic Distillation System for Removing Krypton from Xenon

Instrumentation and Detectors 2022-01-12 v3 High Energy Physics - Experiment

Abstract

An efficient cryogenic distillation system was designed and constructed for PandaX-4T dark matter detector based on the McCabe-Thiele (M-T) method and the conservation of mass and energy. This distillation system is designed to reduce the concentration of krypton in commercial xenon from 5X10710^{-7} mol/mol to 101410^{-14} mol/mol with 99% xenon collection efficiency at a maximum flow rate of 10 kg/h. The offline distillation operation has been completed and 5.75 tons of ultra-high purity xenon was produced, which is used as the detection medium in PandaX-4T detector. The krypton concentration of the product xenon is measured with an upper limit of 8.0 ppt. The stability and purification performance of the cryogenic distillation system are studied by analyzing the experimental data, which is important for theoretical research and distillation operation optimization.