Establishing an independent measurement traceability for 60-Co Air Kerma

Instrumentation and Detectors 2026-07-30 v1 Computational Physics Medical Physics

Abstract

An independent air kerma traceability chain for 60^{60}Co radiation protection levels has been successfully established at the CIEMAT Ionizing Radiation Metrology Laboratory (LMRI-CIEMAT), based on the reference provided by the 137^{137}Cs primary standard. To achieve this, a secondary standard ionization chamber with an appropriate energy response was characterized, and its beam-quality correction factor, kQk_Q, was accurately determined via Monte Carlo simulations using the EGSnrc code. The method's accuracy was validated through a comparison with long-term historical calibration data and peer-confirmed through CIEMAT's successful participation in the EURAMET.RI(I)-S19 supplementary comparison, recently officially published in the BIPM KCDB. The excellent agreement achieved with international reference values demonstrates the high robustness and traceability independence of this newly implemented methodology.

Keywords

Cite

@article{arxiv.2607.27875,
  title  = {Establishing an independent measurement traceability for 60-Co Air Kerma},
  author = {Néstor Cornejo-Díaz and Cristina García-Mulas and Paz Aviés-Lucas},
  journal= {arXiv preprint arXiv:2607.27875},
  year   = {2026}
}

Comments

Accepted for publication in Journal of Instrumentation (JINST reference: JINST_010P_0626)