Dosimetric Study of Lung Modulation and Motion Effects in Carbon ion Therapy for Lung Cancer

Medical Physics 2026-02-18 v1

Abstract

Carbon-ion radiotherapy provides high dose conformity for lung cancer, but its benefit is limited by two sources of uncertainties: interplay between scanned beam delivery and tumor motion, and dose modulation from heterogeneous lung tissue. This study quantifies the separate and combined dosimetric impact of these effects using the GSI TRiP4D treatment planning system. Eighteen lung cancer 4DCT datasets from TCIA were analyzed. A modulation power (PmodP_{\mathrm{mod}}) was assigned to lung voxels. Three values were sampled from a Gaussian distribution (200μm±67μm200\mu\mathrm{m} \pm 67\mu\mathrm{m}), and an extreme value of 750μm750\mu\mathrm{m} was tested. Interplay doses were computed by combining scanned-beam delivery with patient-specific respiratory motion. Four scenarios were studied: static, static with modulation, interplay, and interplay with modulation. Metrics included D95%D95\%, V95%V95\%, homogeneity index (HI), lung V16GyV16\mathrm{Gy}, and heart V20GyV20\mathrm{Gy}. Interplay reduced target coverage by 5.2±1.55.2 \pm 1.5 pp (D95%D95\%), 12.1±5.912.1 \pm 5.9 pp (V95%V95\%), and 8.3±2.48.3 \pm 2.4 pp (HI). Extreme PmodP_{\mathrm{mod}} alone caused small degradations. When combined with interplay, it partially compensated the loss. This effect decreased with 4D optimization. Fractionation mitigated interplay, leaving lung modulation as the main residual effect.

Keywords

Cite

@article{arxiv.2602.15672,
  title  = {Dosimetric Study of Lung Modulation and Motion Effects in Carbon ion Therapy for Lung Cancer},
  author = {Maria Chiara Martire and Lennart Volz and Marco Durante and Christian Graeff},
  journal= {arXiv preprint arXiv:2602.15672},
  year   = {2026}
}