English

Development of wide range photon detection system for muonic X-ray spectroscopy

Instrumentation and Detectors 2023-12-27 v2 Nuclear Experiment

Abstract

We have developed a photon detection system for muonic X-ray spectroscopy. The detector system consists of high-purity germanium detectors with BGO Compton suppressors. The signals from the detectors are readout with a digital acquisition system. The absolute energy accuracy, energy and timing resolutions, photo-peak efficiency, the performance of the Compton suppressor, and high count rate durability are studied with standard γ\gamma-ray sources and in-beam experiment using 27Al(p,γ)28Si^{27}\mathrm{Al}(p, \gamma){}^{28}\mathrm{Si} resonance reaction. The detection system was demonstrated at Paul Scherrer Institute. A calibration method for a photon detector at a muon facility using muonic X-rays of 197^{197}Au and 209^{209}Bi is proposed.

Keywords

Cite

@article{arxiv.2309.00220,
  title  = {Development of wide range photon detection system for muonic X-ray spectroscopy},
  author = {R. Mizuno and M. Niikura and T. Y. Saito and T. Matsuzaki and H. Sakurai and A. Amato and S. Asari and S. Biswas and I. Chiu and L. Gerchow and Z. Guguchia and G. Janka and K. Ninomiya and N. Ritjoho and A. Sato and K. von Schoeler and D. Tomono and K. Terada and C. Wang},
  journal= {arXiv preprint arXiv:2309.00220},
  year   = {2023}
}