Radon-induced backgrounds, particularly from 222Rn and its beta-emitting progeny, present a critical challenge for next-generation liquid xenon (LXe) detectors aimed at probing dark matter down to the neutrino fog. To address this, we developed a compact hermetic gaseous xenon (GXe) detector. This device physically isolates the active volume from external radon sources by using a PTFE vessel sealed between two quartz flanges with mechanically compressed ePTFE gaskets. To quantify radon sealing performance, we implemented a dual-loop GXe circulation system and conducted a 670-hour radon-injection measurement campaign. Radon ingress into the hermetic detector was monitored using electrostatic radon detectors and photomultiplier tubes (PMTs). From these two independent measurements, the steady-state ratios of the radon concentrations inside the hermetic detector to those outside were estimated to be (1.1±0.1)×10−2 and (1.1±0.2)×10−2, corresponding to radon-leakage flows of (2.9±0.3)×10−11 and (2.6±0.4)×10−11m3s−1, respectively. An extrapolation to a 60-tonne LXe TPC such as XLZD suggests that the radon leakage could amount to 1.2×10−2mBq, which is negligible compared to the expected natural radon emanation inside the detector, typically 3mBq. These results demonstrate that flange-based mechanical sealing provides an effective solution for realizing radon-isolated inner detectors in large-scale LXe experiments.
@article{arxiv.2511.03250,
title = {Development of a Hermetic Gaseous Xenon Detector for Suppressing External Radon Background},
author = {Ryuta Miyata and Koki Fujikawa and Rina Harata and Yoshitaka Itow and Shingo Kazama and Masatoshi Kobayashi},
journal= {arXiv preprint arXiv:2511.03250},
year = {2025}
}