English

Development and validation of a cryogenic far-infrared diffraction grating spectrometer used to post-disperse the output from a Fourier transform spectrometer

Instrumentation and Detectors 2024-01-31 v2 Instrumentation and Methods for Astrophysics

Abstract

Recent advances in far-infrared detector technology have led to increases in raw sensitivity of more than an order of magnitude over previous state-of-the-art detectors. With such sensitivity, photon noise becomes the dominant noise component, even when using cryogenically cooled optics, unless a method of restricting the spectral bandpass is employed. The leading instrument concept features reflecting diffraction gratings which post-disperse the light that has been modulated by a polarizing Fourier transform spectrometer (FTS) onto a detector array, thereby reducing the photon noise on each detector. This paper discusses the development of a cryogenic (4 K) diffraction grating spectrometer which operates over the wavelength range from 285 - 500 μ\mum and was used to post-disperse the output from a room-temperature polarizing FTS. Measurements of the grating spectral response and diffraction efficiency are presented as a function of both wavelength and polarization to characterize the instrumental performance.

Keywords

Cite

@article{arxiv.2401.04488,
  title  = {Development and validation of a cryogenic far-infrared diffraction grating spectrometer used to post-disperse the output from a Fourier transform spectrometer},
  author = {Alicia M. Anderson and David A. Naylor and Brad G. Gom and Matthew A. Buchan and Adam J. Christiansen and Ian T. Veenendaal},
  journal= {arXiv preprint arXiv:2401.04488},
  year   = {2024}
}

Comments