Detection of low energy electrons by a total absorption Cherenkov spectrometer

Instrumentation and Detectors 2019-12-02 v1

Abstract

Presented are results of calibration, with a quasi-monochromatic electron beam, of a total absorption Cherenkov spectrometer based on the 14.8 X0 lead glass TF-1. It was found that the energy resolution of the spectrometer was 89% to 10% in the energy range of the electron beam E = 6 to 285 MeV, respectively.

Keywords

Cite

@article{arxiv.1911.12608,
  title  = {Detection of low energy electrons by a total absorption Cherenkov spectrometer},
  author = {V. I. Alekseev and V. A. Baskov and V. A. Dronov and A. I. L'vov and A. V. Koltsov and Yu. F. Krechetov and E. I. Malinovsky and V. V. Polyansky},
  journal= {arXiv preprint arXiv:1911.12608},
  year   = {2019}
}

Comments

6 pages, 5 figures