Detection efficiency evaluation for a large area neutron sensitive microchannel plate detector

Instrumentation and Detectors 2017-02-08 v1

Abstract

In this paper, the detection efficiency of a large area neutron sensitive microchannel plate detector has been evaluated. A 6LiF/ZnS detector was employed as the benchmark detector, the TOF spectra of these two detectors were simultaneously measured and the energy spectra were then deduced to calculate the detection efficiency curve of the nMCP detector. Tests show the detection efficiency@25.3 meV thermal neutron is 34% for this nMCP detector.