Correcting impedance measurements for background parasitics to characterize circuit components in cryogenic environments

Instrumentation and Detectors 2025-06-17 v1

Abstract

Predictable circuit response is a critical prerequisite for accurate electronic measurements. We describe a powerful, yet straightforward, experimental method and analysis model that utilizes an affordable LCR meter in conjunction with an in situ parasitic impedance background correction procedure to measure the temperature-dependent impedance (magnitude and phase) of individual passive circuit elements mounted in a cryostat. We show how the model unambiguously identified a 20x drop in capacitance for 22 microF 5XR multilayer ceramic capacitors cooled from 300 K to 360 mK in an environment with parasitic capacitance of order 300 pF. The same experimental procedure, based on a simple two-wire measurement, was also used to successfully measure 10 pF and 22 pF thin-film capacitors and 100 MOhm thick-film resistors. The results showed that the resistor values increased by up to an order of magnitude when the devices were cooled from 300 K to 360 mK. Most importantly, the simple data acquisition method and robust analysis model were shown to effectively extend the accuracy of a simple benchtop LCR meter beyond its manufacturer-guaranteed values for a wide range of measurement frequencies.

Keywords

Cite

@article{arxiv.2506.12268,
  title  = {Correcting impedance measurements for background parasitics to characterize circuit components in cryogenic environments},
  author = {Riley J. Carpenter and Jadyn Anczarski and Ivar Rydstrom and Aviv Simchony and Zoe J. Smith and Noah A. Kurinsky and Betty A. Young},
  journal= {arXiv preprint arXiv:2506.12268},
  year   = {2025}
}