Characterizing the Experiment for Calibration with Uranium (Excalibur) Neutron Source for Use in Warhead Verification

Instrumentation and Detectors 2025-03-19 v1 Nuclear Experiment

Abstract

Neutron sources can play a variety of roles in warhead verification. For transmission radiography, a source of directed high energy neutrons is required, while for applications to detect fissile isotopes, sub-MeV neutrons are preferred. The Excalibur (Experiment for Calibration with Uranium) neutron source has been built and used in a variety of verification-related experiments. Excalibur is based on a commercial deuterium-tritium neutron generator specified and measured to be capable of producing 14 MeV neutrons at rates of up to 8.2×\times108^8 neutrons/s. The generator is enclosed in a carbon-steel 32^{\prime\prime} diameter, 23.62^{\prime\prime} high carbon-steel cylinder that moderates the mean neutron energy to under 500 keV. This, in turn, is encased in 5\%-borated polyethylene such that the entire assembly is a 48^{\prime\prime}×\times48^{\prime\prime} box that is 30^{\prime\prime} tall. For radiographic applications, a narrow, tapered channel in the steel and polyethylene allows 14 MeV neutrons to stream directly from the generator to a test object. Its collimating capability is demonstrated by measuring the neutron flux profile. In the moderated mode of operation, the generator is fully enclosed in the steel, but a large section of the polyethylene is removed, providing a flux of sub-MeV neutrons from a wide range of angles. Neutron angular and spectral measurements using both a nested neutron spectrometer and a commercial liquid scintillator coupled with a 3^3He detector show the expected softer neutron spectrum in moderated mode in good agreement with MCNP6 calculations. The gamma-ray spectrum from Excalibur is also in good agreement with MCNP modeling. Based on these findings, the future application of Excalibur in its two configurations is discussed.

Keywords

Cite

@article{arxiv.2503.13636,
  title  = {Characterizing the Experiment for Calibration with Uranium (Excalibur) Neutron Source for Use in Warhead Verification},
  author = {Jihye Jeon and Erik P. Gilson and Michael Hepler and Alexander Glaser and Robert J. Goldston},
  journal= {arXiv preprint arXiv:2503.13636},
  year   = {2025}
}
R2 v1 2026-06-28T22:24:18.651Z