English

Calibration of photomultiplier using UVLED

Instrumentation and Detectors 2018-12-05 v1

Abstract

Detector calibration is very important for the long-term operation. For the purposes of simple and precise calibration, we developed a new portable calibration source using UVLED to calibrate the fluorescence detector of the Telescope Array experiment (TA). The UVLED is light (less than 1 kg) and its setup is quick and easy. Therefore, a single source will be used instead of the calibration sources that TA currently uses. The UVLED emits 369 nm wavelength and ~70 pJ of photons which are calibrated with 5% accuracy us-ing photo-diode. It has a particular unit, which is an internal heater, to keep the temperature constant and avoiding temperature dependency on light yield. The first result of calibration for 24 detectors is xx pho-tons/faced count for 369 nm photons with 6% uncertainty. It is in agreement with the result of the current calibration sources of TA.

Keywords

Cite

@article{arxiv.1812.01364,
  title  = {Calibration of photomultiplier using UVLED},
  author = {Sungrae Im and Bokkyun Shin and Ikeda Daisuke and Stan Thomas and ByungGu Cheon},
  journal= {arXiv preprint arXiv:1812.01364},
  year   = {2018}
}