Analysis of Short-Term Stability of Miniature 171Yb+ Buffer Gas Cooled Trapped Ion Clock
Atomic Physics
2020-04-27 v1
Abstract
We demonstrate an improvement in short-term stability by a factor of 10 over a previous generation miniature buffer gas cooled trapped ion clock. We describe the enhancement to detection SNR that has enabled this improvement, the method of clock operation, and the measurement of clock short-term stability. Additionally, we numerically investigate the magnitude of the Dick effect in our pulsed ion clock.
Keywords
Cite
@article{arxiv.1802.04832,
title = {Analysis of Short-Term Stability of Miniature 171Yb+ Buffer Gas Cooled Trapped Ion Clock},
author = {David R. Scherer and C. Daniel Boschen and Jay Noble and Michael Silveira and Dwayne Taylor and Jonathan Tallant and K. Richard Overstreet and S. R. Stein},
journal= {arXiv preprint arXiv:1802.04832},
year = {2020}
}
Comments
submitted to ION Precise Time and Time Interval Meeting, Reston, VA, 2018