Abstract

We have developed a novel in-vivo dosimeter based on glass encapsulated TLD rods. These encapsulated rods can be injected into mice and used for validating doses to an individual mouse in a protracted irradiation scenario where the mouse is free to move in an inhomogenous radiation field. Data from 30 irradiated mice shows a reliable dose reconstruction within 10% of the nominal delivered dose.

Cite

@article{arxiv.1904.07311,
  title  = {An injectable dosimeter for small animal irradiations},
  author = {Guy Garty and Monica C. Pujol and David J. Brenner},
  journal= {arXiv preprint arXiv:1904.07311},
  year   = {2019}
}