English

An In-Situ Movable Calibration Source for Cryogenic Particle Detectors

Instrumentation and Detectors 2020-05-06 v2 High Energy Physics - Experiment

Abstract

A prototype device capable of moving a radioactive calibration source to multiple positions was operated at millikelvin temperatures using a modified commercial stepper motor. It was developed as an in-situ calibration strategy for cryogenic dark matter detectors. Data taken by scanning a calibration source across multiple radial positions of a prototype dark matter detector demonstrated its functionality. Construction, heat load, and operation of the device are discussed, as is the effect of the motor on the detector operation. A sample dataset taken over multiple positions of a SuperCDMS detector is presented as an example of the utility of such a device.

Keywords

Cite

@article{arxiv.2002.00893,
  title  = {An In-Situ Movable Calibration Source for Cryogenic Particle Detectors},
  author = {N. Mast and M. Fritts and D. Sincavage and P. Cushman and V. Mandic},
  journal= {arXiv preprint arXiv:2002.00893},
  year   = {2020}
}

Comments

5 pages, 4 figures

R2 v1 2026-06-23T13:29:35.170Z