Achromatizing a liquid-crystal spectropolarimeter: Retardance vs Stokes-based calibration of HiVIS
Abstract
Astronomical spectropolarimeters can be subject to many sources of systematic error which limit the precision and accuracy of the instrument. We present a calibration method for observing high-resolution polarized spectra using chromatic liquid-crystal variable retarders (LCVRs). These LCVRs allow for polarimetric modulation of the incident light without any moving optics at frequencies >10Hz. We demonstrate a calibration method using pure Stokes input states that enables an achromatization of the system. This Stokes-based deprojection method reproduces input polarization even though highly chromatic instrument effects exist. This process is first demonstrated in a laboratory spectropolarimeter where we characterize the LCVRs and show example deprojections. The process is then implemented the a newly upgraded HiVIS spectropolarimeter on the 3.67m AEOS telescope. The HiVIS spectropolarimeter has also been expanded to include broad-band full-Stokes spectropolarimetry using achromatic wave-plates in addition to the tunable full-Stokes polarimetric mode using LCVRs. These two new polarimetric modes in combination with a new polarimetric calibration unit provide a much more sensitive polarimetric package with greatly reduced systematic error.
Keywords
Cite
@article{arxiv.1002.1409,
title = {Achromatizing a liquid-crystal spectropolarimeter: Retardance vs Stokes-based calibration of HiVIS},
author = {D. M. Harrington and J. R. Kuhn and C. Sennhauser and E. J. Messersmith and R. J. Thornton},
journal= {arXiv preprint arXiv:1002.1409},
year = {2015}
}
Comments
Accepted in PASP