English

Access to the full 3D Brillouin zone with time resolution, using a new tool for pump-probe ARPES

Materials Science 2023-09-22 v1

Abstract

Here we report the first time- and angle-resolved photoemission spectroscopy (TR-ARPES) with the new Fermiologics "FeSuMa" analyzer. The new experimental setup has been commissioned at the Artemis laboratory of the UK Central Laser Facility. We explain here some of the advantages of the FeSuMa for TR-ARPES and discuss how its capabilities relate to those of hemispherical analyzers and momentum microscopes. We have integrated the FeSuMa into an optimized pump-probe beamline that permits photon-energy- (i.e., kz-) dependent scanning, using probe energies generated from high harmonics in a gas jet. The advantages of using the FeSuMa in this situation include the possibility of taking advantage of its "fisheye" mode of operation.

Keywords

Cite

@article{arxiv.2309.11535,
  title  = {Access to the full 3D Brillouin zone with time resolution, using a new tool for pump-probe ARPES},
  author = {Paulina Majchrzak and Yu Zhang and Andrii Kuibarov and Richard Chapman and Adam Wyatt and Emma Springate and Sergey Borisenko and Bernd Büchner and Philip Hofmann and Charlotte E. Sanders},
  journal= {arXiv preprint arXiv:2309.11535},
  year   = {2023}
}

Comments

Submitted to the Review of Scientific Instruments