Absolute Calibration of a Time-Resolved High Resolution X-ray Spectrometer for the National Ignition Facility (invited)
Abstract
A high resolution, Diagnostic Instrument Manipulator (DIM)-based x-ray Bragg crystal spectrometer has been calibrated for and deployed at the National Ignition Facility (NIF) to diagnose plasma conditions in ignition capsules near stagnation times. The spectrometer has two conical crystals in the Hall geometry focusing rays from the Kr He, Ly, and He complexes onto a streak camera, with the physics objectives of measuring time-resolved electron density and temperature through observing Stark broadening and the relative intensities of dielectronic satellites. A third von H\'amos crystal that time-integrates the Kr He, He and intervening energy range provides in-situ calibration for the streak camera signals. The spectrometer has been absolutely calibrated using a microfocus x-ray source, an array of CCD and single-photon-counting detectors, and multiple K- and L-absorption edge filters at the Princeton Plasma Physics Laboratory (PPPL) x-ray laboratory. Measurements of the integrated reflectivity, energy range, and energy resolution for each crystal are discussed. These calibration data provide absolute x-ray signal levels for NIF measurements, enabling precise filter selection and comparisons to simulations.
Keywords
Cite
@article{arxiv.2607.05766,
title = {Absolute Calibration of a Time-Resolved High Resolution X-ray Spectrometer for the National Ignition Facility (invited)},
author = {Lan Gao and B. F. Kraus and K. W. Hill and M. Bitter and P. Efthimion and M. B. Schneider and A. G. MacPhee and D. B. Thorn and J. Kilkenny and J. Ayers and R. Kauffman and H. Chen and D. Nelson},
journal= {arXiv preprint arXiv:2607.05766},
year = {2026}
}