English

Aberration-corrected transmission electron microscopy with Zernike phase plates

Instrumentation and Detectors 2022-05-10 v2 Optics

Abstract

We explore the possibility of applying physical phase plates (PPs) in combination with aberration-corrected transmission electron microscopy. Phase-contrast transfer characteristics are calculated and compared for a thin-film based Zernike PP, a hole-free (HF) or Volta PP and an electrostatic Zach PP, considering their phase-shifting properties in combination with partial spatial coherence. The effect of slightly converging illumination conditions, often used in high-resolution applications, on imaging with PPs is discussed. Experiments with an unheated Zernike PP applied to various nanomaterial specimens and a qualitative analysis clearly demonstrates the general compatibility of PPs and aberration-corrected transmission electron microscopy. Calculations and experiments show the benefits of the approach, among which is a strong phase-contrast enhancement of a large range of spatial frequencies. This allows the simultaneous imaging of atomic-resolution structures and morphological features at the nanometer scale, with maximum phase contrast. The calculations can explain why the HFPP damps contrast transfer at higher spatial frequencies.

Keywords

Cite

@article{arxiv.2202.00960,
  title  = {Aberration-corrected transmission electron microscopy with Zernike phase plates},
  author = {Simon Hettler and Raul Arenal},
  journal= {arXiv preprint arXiv:2202.00960},
  year   = {2022}
}

Comments

Revised version submitted to Ultramicroscopy 21 pages, 14 figures

R2 v1 2026-06-24T09:15:27.891Z