English

A Novel Low-Rank Tensor Method for Undersampling Artifact Removal in Respiratory Motion-Resolved Multi-Echo 3D Cones MRI

Image and Video Processing 2023-05-02 v1 Medical Physics

Abstract

We propose a novel low-rank tensor method for respiratory motion-resolved multi-echo image reconstruction. The key idea is to construct a 3-way image tensor (space ×\times echo ×\times motion state) from the conventional gridding reconstruction of highly undersampled multi-echo k-space raw data, and exploit low-rank tensor structure to separate it from undersampling artifacts. Healthy volunteers and patients with iron overload were recruited and imaged on a 3T clinical MRI system for this study. Results show that our proposed method Successfully reduced severe undersampling artifacts in respiratory motion-state resolved complex source images, as well as subsequent R2* and quantitative susceptibility mapping (QSM). Compared to conventional respiratory motion-resolved compressed sensing (CS) image reconstruction, the proposed method had a reconstruction time at least three times faster, accounting for signal evolution along the echo dimension in the multi-echo data.