Abstract

Most virtual source models (VSM) use beam modeling, with the exception of the patient-dependent secondary collimator (jaw). Unlike other components of the treatment head, the jaw absorbs many photons generated by the bremsstrahlung, which decreases the efficiency of the simulation. In the present study, a new method of beam modeling using a virtual jaw was applied to improve the calculation efficiency of VSM. The results for the percentage depth dose and profile of the virtual jaw VSM calculated in a homogeneous water phantom agreed with the measurement results for the CC13 cylinder type ion chamber within an error rate of 2%, and the 80 to 20% penumbra width agreed with the measurement results within an error of 0.6 mm. Compared with the existing VSM, in which a great number of photons are absorbed, the calculation efficiency of the VSM using the virtual jaw was expected to increase by approximately 67%.

Cite

@article{arxiv.1503.03230,
  title  = {6 MV photon beam modeling for Varian Clinac iX using GEANT4 virtual jaw},
  author = {Byung Yong Kim and Hyung Dong Kim and Dong Ho Kim and Jong Geun Baek and Su Ho Moon and Gwang Won Rho and Jeong Ku Kang and Sung Kyu Kim},
  journal= {arXiv preprint arXiv:1503.03230},
  year   = {2015}
}

Comments

17 pages, 5 figures